Specially designed for the requirements in laboratories, these autoclaves make sterilisation processes safe, accurate, reproducible and validatable. 3 model ranges with different performance.
Systec VX: For all laboratory applications, incl. sophisticated state-of-the-art sterilisation processes. With numerous options for validatable sterilisation.
Systec VE: For basic laboratory applications. With limited options for process optimisation.
Systec VB: For simple process applications. No options available for process optimisation.
Standard Features:- Housing, support frame and pressure vessel made of corrosion-resistant st. steel.
- Temperature and pressure range 140 °C, 4 bar.
- LCD display and fully automatic microprocessor control.
- Code-secured access rights for changing parameters and further safety-relevant intervention.
- Flexible PT-100 temperature sensor.
- Integrated, separate steam generator.
- Number of sterilisation programmes: up to 25 (on request).
- Internal memory for storing up to 500 sterilisation cycles.
- Timer for starting programmes.
- Autofill: automatic demineralised water feed for steam generation.
- Flexible PT-100 temperature sensor.
- Additional temperature sensor in condense exhaust.
- Temperature holding function for liquids after programme finish.
- Special programme for Durham tubes.
- Calculation of F0 value.
- Special programme for waste sterilisation with pulsed heat-up for more efficient air exhaust.
- Water-cooled steam exhaust, thermostatically controlled.
- Programmable automatic door-opening on completion of programme.
- RS-232 and RS-485 interfaces for external data transmission (network-compatible).